11-12
IM 704510-01E
Example when data is saved in Ascii format
Data Size
The size varies depending on the data type as follows.
• Bin format
Data measured in hardware histogram mode
The data size varies depending on the measurement function and other items as
shown in the table below.
Measurement 1
Measurement 2
Measurement 3
Measurement 4
Approx. 152 to 536 KB Approx. 408 to 1561 KB Approx. 280 to 1048 KB Approx. 793 to 3098 KB
Measurement 1: Period, A-to-B time interval (polarity is A
↑
B
↑
, A
↓
B
↑
, A
↑
B
↓
, or A
↓
B
↓
),
pulse width (polarity is
or
)
Measurement 2: A-to-B time interval (polarity is
A B
or
A B
), period, or pulse width
(polarity is
)
Measurement 3: Period A & period B or period A & A-to-B time interval
Measurement 4: Pulse width A & A-to-B time interval, pulse width A & pulse width B
Data measured in time stamp mode or inter-symbol interference analysis mode
• Single measurement function: Approx. 8
×
the number of s approx. 24 KB
• Dual measurement function:
Approx. 16
×
the number of s approx. 24 KB
• Ascii format
The data size varies greatly depending on the sampling mode, measurement function,
sample size, histogram display, and other settings. The maximum size is
approximately 33 M for time stamp mode, single measurement function, and sample
size of 1024000.
Entering a File Name or a Comment
• Enter a file name using up to 8 characters. Enter a comment using up to 25
characters.
• A file name must be assigned. Comments are optional.
• You cannot save to a file name that already exists in the same directory (overwriting
not allowed).
• The file name is common with other saved data, except the extension.
Auto Naming Function
When Auto Naming is turned ON, files with a three digit number from 000 to 499 are
automatically created when saving the data. A common name (up to five characters,
TA720 by default) is attached before the number.
11.5 Saving and Recalling Measured Results