8-5
IM 704510-01E
Detailed Display Setup
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
Setting the X-Axis of the Window
The setting of the following parameters can be changed for each window.
• Label:
Window name (selectable range is 1 to 16)
• Center:
Center value
• Span:
Window span
• Left Marker:
Position of the X1 marker
• Right Marker: Position of the X2 marker
Statistical calculation is performed on the area enclosed by Left Marker and Right
Marker.
For details on the selectable range of each parameter, see section 8.4.
Selecting the Window to Be Displayed (WindowNum)
Select the window to be displayed from below.
[Label]T:
Window set using the Window Parameter Setting dialog box.
All:
Shows the histogram of all windows and a list the statistical values for
each window.
Summation:
Sums the histogram of all windows and displays the result.
When window is set to All display
Displays the top three parameters
whose statistical calculation has
been turned ON using the Calc/Stat
soft key
Note
•
The center and span values can also be entered for each window on the soft key menu that is
displayed when the SCALE key is pressed.
•
The left marker and right marker values can be entered on the soft key menu that is displayed
when the MARKER key is pressed.
8.2 Setting the Multi Window