14-5
IM 704510-01E
Troubleshooting and Maintenance
3
2
1
4
5
6
7
8
9
10
11
12
13
14
14
15
14.3 Displaying the Error Log
Procedure
1.
Press the UTILITY key to display the Utility menu.
2.
Press the Next1/2 soft key to display the 2/2 menu.
3.
Press the Error Load soft key to display the error log.
If there are many messages and they do not fit in the window, turn the rotary knob
to scroll the messages.
Clearing the Error Messages
4.
Press the Clear soft key on the Error Logging menu.
The messages are cleared and the mark at the upper right corner of the screen is
also cleared.
Explanation
When an operation error, a communication error, or a warning occurs, the TA720 creates
an error message list and displays a mark at the upper right corner of the screen. The
details of the message can be viewed by displaying the error log.
Warning mark
You must press the Clear soft key on the Error Logging menu and clear the message list
to clear the mark from the screen.
Note
Warning messages cannot be cleared using the Clear soft key. Warning messages are
automatically cleared when the cause of the warning is resolved such as by restarting the
measurement.