10-1
IM 704510-01E
Storing Setup Information to the Internal Memory and Recalling Setup Information
3
2
1
4
5
6
7
8
9
10
10
11
12
13
14
15
Chapter 10 Storing Setup Information to the Internal Memory and Recalling Setup Information
10.1 Storing the Setup Parameters to the Internal
Memory
For a functional description, see 2.8.
Procedure
1.
Press the UTILITY key to display the Utility menu.
2.
Press the Memory soft key to display the Memory menu.
Selecting the Memory Number
3.
Turn the rotary knob to select the memory number.
Setting a Comment
4.
Press the Comment soft key to display the keyboard, and enter the characters for
the comment.
By default, MEM + a 2-digit memory number is specified as in MEM05.
For the procedure to set the character string, see section 4.4.
Executing the Store Operation
5.
Press the Store soft key to execute the store operation.
Explanation
Settings That Are Stored
Measurement conditions and display parameters that are set using the following keys
are stored.
MODE key, FUNCTION key, SAMPLE key, INPUT key, DISPLAY key, SCALE key, and
MARKER key
Selecting the Memory Number
Select from the following 32 memory numbers. If the setup parameters are already
stored at the selected number, the previous data is overwritten.
0 to 31
Setting a Comment
The maximum number of characters for a comment is 10. A comment can be entered
for each memory number.
Note
•
The stored setup parameters are not cleared even if you initialize the settings on the TA720.
•
For details on parameters that are not stored, see page 13-3.