15-10
IM 704510-01E
15.14 PC Card Drive (Optional)
Item
Specifications
Number of slots
1
Supported cards
Flash ATA memory card (PC card TYPE II)
15.15 Ethernet Communications (Optional)
Item
Specifications
Communication port
1
Electrical and mechanical
Conforms to IEEE802.3
specifications
Transmission system
Ethernet (10BASE-T)
Transmission rate
10 M bps
Communication protocol
TCP/IP
Supported services
FTP server, FTP client (network drive), DHCP, DNS
Connector type
RJ-45 connector
15.16 General Specifications
Item
Specifications
Electrical and mechanical
• Ambient temperature: 23
±
5
°
C
specifications
• Ambient humidity: 50
±
10% RH
• Supply voltage/frequency error: Within 1% of rating
Operating altitude
2,000 m or less
Warming up
Approx. 30 minutes
Storage conditions
• Temperature: –20 to 60
°
C
• Humidity: 20 to 80%RH (no condensation)
Operating conditions
• Temperature: 5 to 40
°
C
• Humidity: 20 to 80%RH (no condensation)
Rated supply voltage
100 to 120 VAC, 200 to 240 VAC
Permitted supply voltage
90 to 132 VAC, 180 to 264 VAC
range
Rated supply voltage
50/60 Hz
frequency
Permitted supply voltage
48 to 63Hz
frequency range
Maximum power
250 VA
consumption
Withstanding voltage
1.5 kVAC, 10 mA or less for one minute (between power supply and case)
Insulation resistance
500 VDC, 10 M
Ω
or more (between power supply and case)
Signal ground
The ground of all input and output connectors is connected to the case ground.
External dimensions
Approx. 426 (W)
×
177 (H)
×
300 (D) mm (projections excluded)
Weight
Approx. 12 kg (main unit only)
Cooling method
Forced air cooling
Installation position
Horizontal (stacking prohibited)
Battery backup
Setup parameters and time are backed up using the internal lithium battery.
Fuse
*1
Maximum rated voltage: 250 V, maximum rated current: 3.15 A, type: time lag, standard: UL/VDE certified
Part number: A1351EF
Standard accessories
Power cord (1), rubber feet (4), printer roll paper, user’s manual (this manual) (1), communication
interface user’s manual (1)
*1 There is also another fuse inside the unit, but the user cannot replace it. If you believe the fuse inside the unit is blown,
contact your nearest YOKOGAWA dealer as listed on the back cover of this manual.