5-5
IM 704510-01E
Setting the Sampling Mode and Measurement Function
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
5.4
Setting the Pulse Width Measurement
For a functional description, see 2.4.
Procedure
1.
Press the FUNCTION key to display the Function menu.
2.
Press the Function soft key to display the Function selection menu.
3.
Press the Pulse Width soft key.
Selecting the Measurement Channel
4.
Press the Channel soft key to select ChA or ChB.
Selecting the Polarity
5.
Press the Polarity soft key to display the Polarity selection menu.
6.
Press the soft key corresponding to the desired polarity.
Explanation
Selecting the Measurement Channel
Select CH A or CH B.
Selecting the Polarity
Select from the following:
•
: Measures the positive side (from the rising edge to the next falling edge).
•
: Measures the negative side (from the falling edge to the next rising edge).
•
: Measures the positive side and negative side alternately.
Measurement Range of Pulse Width
Varies depending on the sampling mode.
• When in time stamp mode:
6 ns to 20 ms
• When in hardware histogram mode:
6 ns to 3.2
µ
s
• When in inter-symbol interference analysis mode: 10 ns to 3.2
µ
s
Continuous Measurement Conditions
When the measurement edge interval is greater than or equal to 12.5 ns, measurements
are made continuously.
S1
S2
S3
S1
S2
S3
12.5 ns or more
Continuous
measurement
Non-continuous
measurement
12.5 ns or less
When measuring pulse width ( )