Operation
Yamatake Corporation
3-44
Model HGC303 - Heat Value Gas Chromatograph
Bold:
Valid response factor
Semi Auto calibration
The semi auto calibration function will perform auto calibration sequence triggered by
pressing [start] button operation. The Hold time setting in auto calibration box is used.
The next section describes each function in the calibration panel
Table 3-19 Example of calibration action
No
Calibration
OLD RF
NEW RF
Judgement
Signal to
model
HMU303
1
Auto
RF0
RF1
OK
OK
2
Auto
RF1
RF2
OK
OK
3
Auto
RF2
RF3
NG
NG
4
Manual
RF2
RF4
OK
NG=>OK
5
Auto
RF4
RF5
OK
OK
6
Auto
RF5
RF6
OK
OK
7
Manual
RF6
RF7
OK
OK
8
Auto
RF7
RF8
OK
OK
CAUTION
(1) For manual calibration and semi auto calibration, the RF will be the new RF
even if the |RF% DEV| > “%DEV RF Limit”. So the judgement is always OK.
(Example No.4 and 7)
(2) Only the checked components in calibration box are recognized and judged.
Refer to No.11 on “Table 3-20 Calibration factor setting” on page 3-46 for
details on the check box of each component.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...