Operation
Yamatake Corporation
3-2
Model HGC303 - Heat Value Gas Chromatograph
3-1-3 : Power on
Supply the power to operate the model HGC303 system according to the following
action.
~Note
After turning on the power, allow 2 hours for the device to warm up.
The carrier gas pressure SP and oven temperature SP have already been factory set in
the model HGC303, therefore, the user doesn't have to worry about setting this data.
Carrier gas pressure SP: less than 300 kPa (43.5 psi)
(SP differs with each model HGC303)
Oven temperature SP: 58
°
C (136.4
°
F)
Analyzing cycle: 300 sec.
~Note
When the power is supplied to the model HGC303, a model HGC303
status error will appear on HGM monitoring system (oven temperature
error message etc.
This is because of a self-diagnostic system error, not a model HGC303
system error.
The model HGC303 status will automatically return to normal once the
oven temperature reaches 58
°
C (136.4
°
F).
Table 3-2 The procedure to start up the model HGC303 system
Step
Action
1
Supply the valve operating gas
2
Supply the carrier gas pressure
3
Supply the power to the model HGC303
4
Supply the power to the model HMU303
5
Wait until the model HGC303 system becomes stale.
6
Supply the process gas
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...