Installation
Yamatake Corporation
2-20
Model HGC303 - Heat Value Gas Chromatograph
2-5-5 : Model HGC303 wiring
Remove the terminal cover and wiring while referring to the figure and table below.
Figure 2-12 Wiring location
Either internal grounding (earthing) terminal (A) or external grounding (earthing) ter-
minal (B) can be used.
At least one grounding (earthing) terminal connection is recommended.
~Note
Yamatake recommends cable of conductor cross-sectional area 2 (mm
²
) or
equivalent for power supply connection and GND connection.
Table 2-5: Wiring description
Terminal No.
Description
1
Power supply (-)
2
Power supply (+)
3
No connection
4
FB terminal (-)
5
FB terminal (+)
6
No connection
7
Terminator (-)
8
Terminator (+)
A
Internal GND
B
External GND