Yamatake Corporation
Safety
Model HGC303 - Heat Value Gas Chromatograph
vii
4. Class II, Division 2 locations
4.1
Wiring methods
•
Rigid metal conduit, intermediate metal conduit, electrical metallic tubing,
dust-tight wireways, or Type MC or MI cable with approved termination
fittings
, or
Type PLTC in cable trays
, or
Type ITC in cable trays
, or
Type
MC
or
TC cable installed in ladder, ventilated trough, or ventilated channel
cable trays in a single layer, with a space not less than the larger cable
diameter between the two adjacent cables
, can be employed.
• All boxes and fittings must be dusttight.
4.2
Sealing
• Sealing means must be provided as shown in 3.2.
5. Class III, Division 1 locations
5.1
Wiring methods
•
Rigid metal conduit, rigid non-metallic conduit, intermediate metal conduit,
electrical metallic tubing, dust-tight wireways
, or
Type MC or MI cable with
approved termination fittings
, can be employed.
• All boxes and fittings must be dusttight.
5.2
Sealing
• Sealing means are not required.
6. Class III, Division 2 locations
6.1
Wiring methods
• Wiring methods must comply with 5.1.
6.2
Sealing
• Sealing means are not required.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...