Yamatake Corporation
Operation
Model HGC303 - Heat Value Gas Chromatograph
3-23
3-3-6 : Main displays of HGM
Indication panel
Data is updated every 5 minutes.
Table 3-11 Description of the indication panel
No.
Panel
Description
1
Date time
Present date and time
2
Status
Communication status appears when
HGM is communicating with model
HGC303.
3
Data box
Select a data type.
Default: Retention time (sec.)
4
Calculated Value Select values for Ideal gas or Real gas
Default: Real
SCV: Superior Calorific Value
ICV: Inferior Calorific Value
5
Oven Temp. and
Carr. Pres.
Display oven temperature and carrier gas
pressure
6
Chromatogram
The last 300 chromatograms are stored in
RAM. Save the data as required. Select
[previous XX] or [latest] to view the chro-
matogram.
If [previous XX] is selected, the auto
reload function stops. XX: 01-299
Return to [latest] to monitor the latest
chromatogram.
Auto reload function starts again.
7
[Field work]
Model HGC303 holds outputs to the host
control system during field maintenance.
Click on [Field work] then [ON], to set
the holding time to [24hrs]. [Field work]
button blinks while performing fieldwork.
8
[Calibration]
Click on the [Calibration] button to per-
form calibration. The [Calibration] but-
ton blinks during auto calibration.
Refer to “3-4 : Calibration” on page 3-40
9
[Report]
Click on the [Report] button to create a
report.
Refer to “3-3-7 : Report” on page 3-28.
10
HGC Status
Green means that model HGC303 is ana-
lyzing normally.
If this signal changes to red, click on this
button to read the error message.
Refer to “Chapter 5 : Troubleshooting” on
11
Return Menu
Exit from User's mode
Return to Main Menu
1
2
3
4
5
6
7
8
9
10
11
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...