Model HGC303 - Heat Value Gas Chromatograph
5-1
Chapter 5 : Troubleshooting
You will need to troubleshoot if you observe a red error sign flashing on the HGM
screen while the model HGC303 is running.
Click on the error box on the HGM screen to get more detailed information regarding
the error.
Table 5-1 shows the self-diagnostic functions of the model HGC303.
Table 5-1 Model HGC303 self-diagnostics
Error
message
Error description
Action
Oven
temperature
extremely high
Temperature sensor
failure
This is a critical problem. Turn off the power
of HGC. Contact a closer Yamatake office or
distributor.
Carrier Gas
pressure error
Carrier gas pressure
PV
PV<50kPa
or
300kPa<PV
1. Check the carrier gas supply pressure
setting (400+/-50kPa)
2. If the carrier gas PV is equal with carrier
gas supply pressure, this is a critical
problem. Turn off the power and stop the
carrier gas supply.
Contact a closer Yamatake office or
distributor.
Response
Factor error
Auto calibration fail.
RF percent deviation
limit< RF% DEV
Check the standard gas connection,
composition and supply.
This error continues until the correct
calibration take place.
Chromatogram
baseline error
TCD baseline out of
range
1. Check whether the carrier gas type and
purity is correct.
2. If the problem is not caused by the carrier
gas, a critical problem may have occurred.
Oven
temperature
error
Oven temperature
not stable
SP+/-2°C<PV
Verify whether a sudden change has occurred
in the ambient temperature. Shield the HGC
to resume normal operation.
**
Carrier gas
pressure out of
control
Carrier gas pressure
SP+/-30kPa<PV
1. Check the carrier gas supply pressure
setting (400+/-50kPa)
2. Check the remaining gas cylinder
pressure.
3. If the problem is not caused by the carrier
gas, a critical problem may have occurred.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...