Operation
Yamatake Corporation
3-24
Model HGC303 - Heat Value Gas Chromatograph
Heat value and total raw concentration
This graph shows the heat value and the total raw concentration.
The left vertical axis represents the heat value and the right vertical axis represents the
total raw concentration. The horizontal axis represents the time range.
Figure 3-8 Trend graph of SCV and the total concentration (Raw)
* In case of Windows 98 or Me, data are also saved in C:\My documents\MyHGC.
Table 3-12 Trend graph of SCV and total raw concentration description
No.
Display
Description
1
URV (SCV)
Upper Range Value for SCV, default value: 42 MJ/m
3
Click on the URV button to change the URV value
2
Load (Offline)
Recall saved data
File name extension: .hv1
3
Save (Online)
The latest data is saved
Default directory is “C:\Program files\hgm\data”.*
4
Verify that your printer is connected and working properly.
5
30min.
(The time range
select)
This indicates the time range of the horizontal axis.
Select a time range from the pull-down menu: 30 min., 60 min., 3 hour, 6
hours, 12 hours, 1 day, 2 days, 3 days, 6 days, 12 days
Default: 30 min.
6
LRV(SCV)
Lower Range Value for SCV, default value: 38 MJ/m
3
Click on the LRV button to change the LRV value
7
URV
(Total raw conc.)
Upper Range Value for Total, default value: 103%
Click on the URV button to change the URV value
8
LRV
(Total raw conc.)
Lower Range Value for Total, default value: 97%
Click on the LRV button to change the LRV value
9
Black circle
Black indicates SCV graph.
Click on the [Disp data] button (No.11) to select a data type.
Default: SCV (Real)
10 Red circle
Red indicates Total raw conc. graph.
11 Disp. data
Select a data type for heat value.
12 Time
Online: Date and time of the latest data (data is reloaded every 5 min.)
Offline: Date and time of when the data was saved.
HGM000016000S
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...