Safety
Yamatake Corporation
iv
Model HGC303 - Heat Value Gas Chromatograph
2. Cable systems
2.1
Thermoplastic sheathed cables, thermosetting sheathed cables
, or
elastomeric
sheathed cables
can be selected for fixed wiring in the hazardous areas.
2.2 Flameproof cable entry devices (cable glands) certified to comply with EN 50018
and appropriate to the type of cable employed, must be used for the connection of
cables to the apparatus.
3. Conduit systems
For conduit systems, relevant national standards or codes of practice are followed
prior to the following recommendations.
3.1
Screwed heavy gauge steel, solid drawn or seam welded conduit
, or
flexible
conduit for protection of cables in explosive atmospheres
(see ISO 10807) can
be selected for fixed wiring in the hazardous areas.
3.2
Conduit must be threaded for connection
to permit the full engagement of five
threads.
3.3 Either
conduit entry devices
or
sealing devices such as stopping boxes
are
provided at the wall of the apparatus enclosure to limit the pressure piling effect
and to prevent hot gases from entering the conduit system from the enclosure
containing a source of ignition.
Each type of both the devices must be certified
to comply with EN 50018.
3.4
The stopping boxes
, if used,
are filled with a compound
which does not shrink
or setting and is impervious to, and unaffected by, chemicals found in the
hazardous area.
The depth of the compound in the stopping box
is at least equal
to the internal diameter of the conduit, but in no case less than 10 mm.
3.5 When the conduit contains three or more
non-seathed single or multi-core
cables
, the total cross-sectional areas of cables, including insulation, are not more
than 40% of the cross-sectional area of the conduit.
4. Installation in explosive atmospheres caused by air / dust mixtures
4.1
Conduit
or
cable glands
, if employed to connect cables to the apparatus, must be
selected and used in such a way that an
IP6X protection
(dust-tight) is guaranteed.
4.2 It is recommended to maintain the apparatus so that
the dust layer will not exceed
a thickness of 5 mm
.
~Note
Where
the ignition temperature of a dust layer up to 5 mm thickness
is
equal to, or higher than, the value that is obtained by adding
75K
to the
maximum surface temperature of the enclosure “T...
°
C” as marked on the
apparatus, the apparatus is incapable of causing ignition of the dust
layer. (T...
°
C is based on the maximum ambient temperature)
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...