Yamatake Corporation
Operation
Model HGC303 - Heat Value Gas Chromatograph
3-17
Table 3-7 Set up online mode
Step
Action
1
[Analyzer Status]
Select [HGC] in Analyzer Status
If [HGC] cannot be selected from the pull-down menu, click on [Refresh].
The HGM searches for the model HGC303 again along the Fieldbus line.
2
[Data saving interval]
Select “data saving interval” from pull-down menu;
5 min. / 5 sec. [Default]
10 min. / 10 sec.
15 min. / 15 sec.
30 min. / 30 sec.
60 min. / 60 sec.
5 min.: Heat value and Total (Raw) data (text file extension:.hv1)
5 sec.: Oven temperature and Carrier gas pressure data (text file exten-
sion:.sv1)
Refer to “ Data save” on page 3-18 and “ Editing data” on page 3-19 for
details on how to save and edit the data.
3
[Auto saving interval]
Check the box to select an interval as required.
Selection: Min. 1 day, Max 10 day
Refer to “ Automatic file saving” on page 3-20 for details on the auto saving
mechanism.
4
[Calculation Mode]
Select [ISO] or [GPA] from Calculation Mode.
ISO [Default]
5
[Password]
Some screens require a password to access them.
However, if you want to change a password, click on [Change password>>].
The password-setting screen appears on the setup HGM display (See “Figure
3-5 Set up HGM display” on page 3-15). Click on the specified button, and
then enter the “Old password”, which has been stored in the HGM and then
enter a “New password”.
The new password becomes active once you click on [OK] in the password-
setting screen.
Default passwords are as follows (Maximum letters: 16):
Calibration
:
password1
Configuration mode
:
password2
Maintenance mode
:
password3
Field work
:
password4
Calibration data change*: password5
~Note
*This refers to the [Advanced>>] button in “Figure 3-15 Cali-
bration setting panel” on page 3-45.
6
If necessary, click on [Extended setup], and select normalization method.
Default is “Standard normalization”.
7
Click on [OK] to return to the main menu.
8
Click on [User's mode] in the main menu.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...