Installation
Yamatake Corporation
2-4
Model HGC303 - Heat Value Gas Chromatograph
2-2 : Selecting an interface and installation
The HGM requires a data converter to convert the data that is being received from the
HGC. You may use either 1) the Yamatake Heat Value Gas Chromatograph Fieldbus
Adaptor (model HFA100) or 2) the National Instruments Fieldbus-PC Interface.
~Note
1. A PC with a USB1.0 port is required to run the model HFA100.
2. For the National Instruments Fieldbus-PC Interface, you additionally
have to prepare a floppy disk drive and a PC having either a type-II
PCMCIA socket or a 16-bit ISA slot.
2-2-1 : Installing the model HFA100
Installing the hardware:
Refer to the model HFA100 instruction manual.
Installing the software:
In order to use the model HFA100, following two softwares must be installed.
1. HFA Config.exe (Installed at the same time the HGM is installed)
2. HFBADRV (An USB driver for model HFA100)
Procedure for software installation
(1) First install the HGM by following instructions found in section 2-3 “HGM Instal-
lation” below.
(2) Once HGM installation has completed, install 'HFBADRV'.
When connecting the model HFA100 to a USB port, an installation wizard will
appear on your screen. Follow the instructions that the installation wizards give
you. You will be asked for your CD-ROM that came with the model HGC303
where you should select the folder “HFBADRV”. (Once the file ‘HFBADRV’ is
installed, cannot be uninstalled.)
CAUTION
Model HFA100 operates only on Windows2000 / XP. Purchase the National Instru-
ments Fieldbus-PC Interface if you are running an operating system other than Win-
dows 2000/XP.
CAUTION
An installation CD-ROM comes supplied with your model HGC303. Take note that
there are two types of “Setup HGM” folders on the CD-ROM: (1) Setup HGM for
HFA and (2) Setup HGM for NI.
To use the model HFA100, you must select the {setup} icon that is in the “Setup
HGM for HFA” folder. If you by chance install from the “Setup HGM for NI”
folder, uninstall it, and then install from the “Setup HGM for HFA” folder.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...