Operation
Yamatake Corporation
3-48
Model HGC303 - Heat Value Gas Chromatograph
3-4-5 : Report
Click on [Preview] and the following preview screen will appear.
Figure 3-17 Preview screen of report
RF% DEV
The result of RF% DEV is displayed.
Refer to “ RF% DEV (Response Factor Percent Devitation)” on page 3-43 for more
details.
%DEV RF Limit
The allowable tolerance of “RF% DEV” is displayed.
Default value is 10% (absolute value)
These values can be changed only in Maintenance mode.
Judgement
|RF% DEV|
≤
“%DEV RF Limit” => “OK“
|RF% DEV|
>
“%DEV RF Limit” => “NG“
If all judgements were found to be “OK“, the checked components will be calibrated and
the New RF will be used as the response factor.
If an NG message appears, no components will be calibrated and the Old RF will be used
as the response factor and the model HGC303 will transmit an “RF error” to the host
control system via the model HMU303.
Load
Recalls saved data. File name extension: .cdt
Save
The latest component data table is saved.
Default directory is C:\program files\hgm\data
Preview
Latest response factor data is displayed.
The report can also be printed out.
Calibration start
For manual calibration.
This button only appears when in Manual calibration mode.
Refresh
HGM recalculates “new RF” and “RF%DEV”.
This button only appears when in Manual calibration mode.
Table 3-21 Description of component data table
Item
Description
Report Date
Time
File Date
Filename
Time
COMP
CAL
NAME
CONC
1 C6+
0.100
3413.20
34186.56
34234.71
0.14
OK
2 C3H8
1.510
6441.72
4247.64
4266.04
0.43
OK
3 i-C4H10
0.309
6399.25
20435.53
20709.54
1.34
OK
4 n-C4H10
0.298
5289.30
17698.05
17749.33
0.29
OK
5 neo-C5H12
0.100
1726.15
17286.04
17330.82
0.26
OK
6 i-C5H12
0.122
1547.22
12834.21
12682.13
-1.18
OK
7 n-C5H12
0.103
1160.49
11066.80
11266.93
1.81
OK
8 N2
2.980
2112.48
725.84
708.89
-2.34
OK
9 CH4
90.966
29580.98
324.99
325.19
0.06
OK
10 CO2
0.503
3717.20
7411.53
7390.06
-0.29
OK
11 C2H6
3.010
4480.58
1485.46
1488.57
0.21
OK
13:23:30
JUDGE
RF% DEV
PV
NEW PH
OLD RF
NEW RF
01120701.cdt
Dec-07-2001
Dec-07-2001
18:30:00
HGC Calibration
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...