Operation
Yamatake Corporation
3-50
Model HGC303 - Heat Value Gas Chromatograph
Auto calibration
If you need to cancel auto calibration, click on [Stop] from the Auto calibration setting
panel.
Table 3-23 Operating auto calibration procedure
Step
Action
1
Set the model HGC303 Time Adjusting to the present time.
2
Select [Auto calibration].
3
Choose a [Time], [Interval], [Hold time]
(Refer to “Table 3-20 Calibration factor setting” on page 3-46.)
4
Recall the saved calibration data or input concentration of each compo-
nent manually, then check the small box.
5
Click on [Start].
(When Auto calibration is “ON” [STOP] button will turn on, when Auto
calibration is “OFF”, [START] button will turn on.)