Installation
Model HGC303 - Heat Value Gas Chromatograph
2-3
Storing the model HGC303
The model HGC303 should be stored:
- indoor at storage temperature (-40 to 70
°
C); humidity (up to 95%RH)
- in a place safe from vibration or shock.
- in the same packing as it was shipped in.
Model HGC303 that has been used should be stored by following procedures below.
Step
Action
1
Makes sure no process gas remains in the model HGC303.
2
Purge the model HGC303 with helium gas.
3
Insert metal plugs into all the inlets and outlets for carrier
gas, valve operating gas and process gas except VENT
(valve operating gas outlet) in order to keep moisture out.
4
Pack it as it was when it was originally received.
5
Store the model HGC303 indoors at normal temperature
and humidity in a place safe from vibration or shock.