Operation
Yamatake Corporation
3-32
Model HGC303 - Heat Value Gas Chromatograph
Table 3-16 Description of configuration mode display
Display
Description
Outputs configuration
Low cut off: Threshold of a peak or a noise
When detected value < low cut off value, output is transmitted as 0
mol%.
% DEV RF limit: See “3-4-2 : Calibration procedure” on page 3-
Each RF % dev limit can be set independently.
PV12 to 15, 19 and 20 can be configured to set which outputs are
transmitted to the host control system.
PV High / Low Alarm Configuration:
High / Low Alarm limit values can be entered for PV1-11, PV12-
15, PV19-20.
This setting is available for application of
HGC (version 3.1 or later),
HGM (version 4.70 or later)
and HDM (version 2.40 or later) combination.
Reference conditions
Configuration data for heat value calculation can be selected or
entered manually.
Total of raw conc.
Usually, the total of raw concentration is within 95-105 mol% dur-
ing process gas analysis.
If these values are required to be changed, input a user defined
value for both high and low limit.
Normalization method
Normalization method can be selected.
Default is “Standard normalization”
Multi-stream function setting Use or Not use of multi-stream function can be selected.This set-
ting is only available for application of HGC (version 3.0 or later)
and HDM (version 2.30 or later) combination.
Download to HGC
All configured data are downloaded to the model HGC303 by
clicking on this button.
Return Menu
Exit from configuration mode.
Return to main menu.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...