Operation
Yamatake Corporation
3-54
Model HGC303 - Heat Value Gas Chromatograph
3-5-5 : Configuration mode
Numerous configurations of the model HGC303 and the HGM can be done in this mode.
Click on [Configuration mode] from the main menu. The following screen appears
after entering the password.
Figure 3-21 Configuration mode display (GPA)
Table 3-25 Description of configuration mode display (GPA)
Display
Description
Outputs configuration
Low cut off: Threshold of a peak or a noise
When the detected value < low cut off value, the output is
transmitted as 0 mol%.
%DEV RF Limit: See 3-4-2 Calibration procedure on page
3-20 for details.
Each RF %dev limit can be set independently.
PV12 to 15, 19 and 20 can be configured to set which out-
puts will be transmitted to the host control system.
PV High / Low Alarm Configuration:
High / Low Alarm limit values can be entered for PV1-11,
PV12-15, PV19-20.
This setting is available for application of
HGC (version 3.1 or later),
HGM (version 4.70 or later)
and HDM (version 2.40 or later) combination.
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...