Yamatake Corporation
Installation
Model HGC303 - Heat Value Gas Chromatograph
2-19
Prepare the carrier gas and valve operating gas as specified in the table below.
WARNING
Purge the carrier gas line before performing any piping, and then verify that there is
no dust remaining in the piping.
Release the gas from the vent line to the air through the header.
There is a possibility that back-pressure from vent line has a lot of influence.
Table 2-4: Gas specifications
Gas type
Purity
Secondary supply
pressure
Carrier gas
Helium
99.99% or higher
400 ± 50 kPa
(58 ± 7 psi)
Valve operating
gas
Helium, Air,
Nitrogen
99.99% or higher
400 ± 50 kPa
(58 ± 7 psi)
Process gas
Natural gas
-
50 - 490 kPa (7 - 71 psi)
at flow meter inlet