Yamatake Corporation
Operation
Model HGC303 - Heat Value Gas Chromatograph
3-49
3-4-6 : Calibration methods
The model HGC303 has two-calibration methods, manual and automatic.
Manual calibration
Both the [Refresh] and the [Calibration Start] buttons are active when in manual
calibration mode.
Table 3-22 Manual calibration procedure
Step
Action
1
Select [Manual calibration].
2
Recall the saved calibration data or input concentration of each com-
ponent manually.
3
Perform calibration by following the flow chart below.
CAUTION
Perform calibration after verifying that the carrier pressure and oven temperature
are both stable.
Do not input zero into any box.
Do not calibrate without calibration gas.
Data save
Calibration setting panel
End
Component data table
HGC is calibrated
Data load
Print out
Each RF is calculated by HGM
All judgements are OK.
Calibration setting panel
RF error
message will
appear
Component
Data table
Select [Start]
RF%DEV<=%DEV RF limit
Click on [Refresh]
RF%DEV>%DEV RF limit
Click on [Calibration start]
Click on [Close]
Click on [Close]
HGC is NOT calibrated
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...