Yamatake Corporation
Operation
Model HGC303 - Heat Value Gas Chromatograph
3-9
Starting up the HGM with PCMCIA-FBUS
The procedure to start the HGM up are given below.
1. Make sure that both the model HGC303 and the model HMU303 are running
normally.
2. Prepare a personal computer, which has the HGM installed.
3. Make sure the computer is running Windows 98 / Me / NT / 2000 / XP.
4. Verify that font size is [Small font] and the display resolution 1024 × 768
pixels.
5. Connect the model HMU 303 and NIFB-PC interface (Refer to Figure 3-3).
6. Insert the NIFB-PC interface into the appropriate card insertion slot.
7. Make sure that NIFB-PC interface configuration is correct as follows:
(1) Start [Interface Configuration Utility] in NI-FBUS holder.
(2) Select Port 0 and click [Edit].
(3) Select Device Address [Fixed]
and set [0x11].
(4) Select Device Type [Basic Device].
(5) Select Usage [NI-FBUS].