Operation
Yamatake Corporation
3-20
Model HGC303 - Heat Value Gas Chromatograph
Automatic file saving
The HGM can be set to automatically save data files. This is done by activating the
setting from the [setup HGM] panel.
Default directory; C:\program files\hgm\data.
Files with the extensions;.hv1,.cv1, and .sv1 and .cg1(chromatogram) are saved.
All .cg1 files are saved as named YYYYMMDDHHMMas.cg1.
YYYY = year, MM = month, DD = date, HH = hour, MM = minute, as = auto saving,
.cg1 = chromatogram extension file.
Data saving interval of .cg1 files is fixed to 5minutes.
Example:
Auto saving interval:1day (Selection: min. 1day, max. 10 days)
Data saving interval:5 min. and 5 sec. (Selection: min. 5 min. and 5 sec., max 60 min.
and 60 sec.)
(1) HGM data saving starts at 2001/07/25 19:00.
(This function starts after checking the box in [Setup HGM] then clicking on
[OK].)
(2) Analysis data (2001/07/25 19:00-2001/07/25 23:59) is saved at 2001/07/26 0:00
Saved file names: 010725as.hv1, 010725as.cv1, 010725as.sv1.
“as” stands for auto saving.
(3) Analysis data (2001/07/26 0:00-2001/07/26 23:59) is saved at 2001/07/27 0:00.
Saved file names: 010726as.hv1, 010726as.cv1, 010726as.sv1
HGM000012000D
Содержание HGC303
Страница 10: ...Safety Yamatake Corporation viii Model HGC303 Heat Value Gas Chromatograph ...
Страница 42: ...Installation Yamatake Corporation 2 22 Model HGC303 Heat Value Gas Chromatograph ...
Страница 109: ...Yamatake Corporation Operation Model HGC303 Heat Value Gas Chromatograph 3 67 Figure 3 28 User report GPA mode ...
Страница 110: ...Operation Yamatake Corporation 3 68 Model HGC303 Heat Value Gas Chromatograph ...
Страница 112: ...Maintenance Yamatake Corporation 4 2 Model HGC303 Heat Value Gas Chromatograph ...
Страница 127: ...Yamatake Corporation Appendix Model HGC303 Heat Value Gas Chromatograph A 13 Drawings HGC00 ...
Страница 128: ...Appendix Yamatake Corporation A 14 Model HGC303 Heat Value Gas Chromatograph ...