Phadia 100 - User Manual
Routine Assay Run
12-3501-20/12
Page 2.17
2
3
4
5
6
2
13
• ImmunoCAP Specific IgE Negative Control tested for the same three
allergens as above, in duplicate
• At least six individual patient samples in duplicate, any allergens
Criteria for acceptance of Run 1:
• The instrument software accepts the curve
• The mean values for ImmunoCAP Specific IgE Control is within given range
• ImmunoCAP Specific IgE Negative Control will give results below 0.35 kU
A
/l
with allergen ImmunoCAP for all replicates
Run 2 and 3:
• Curve controls 1 and 2 (single replicate, as default)
• ImmunoCAP Specific IgE Control tested for three allergens in duplicate.
• ImmunoCAP Specific IgE Negative Control tested for the same three
allergens as above, in duplicate
• At least six individual patient samples (same as in the first run) in duplicate,
same allergens as in the first run
Criteria for acceptance of Runs 2 and 3:
• Curve controls are within range, the run is accepted by the instrument
software
• The mean values for ImmunoCAP Specific IgE Control are within given
ranges
• ImmunoCAP Specific IgE Negative Control will give results below 0.35 kU
A
/l
with allergen ImmunoCAP for all replicates
Criteria for total acceptance of ImmunoCAP Specific IgE 0-100:
• Criteria for Runs 1 through 3 are fulfilled
• Pooled coefficients of variation (CV) within runs are < 6% for ImmunoCAP
Specific IgE Control and patient samples
• Total CV (%) for ImmunoCAP Specific IgE Control and patient samples are
< 10%
Printouts of laboratory reports and patient reports can be done.
Calculate the CV as described in Calculation of CV % values.
If all the criteria for total acceptance are fulfilled, results can be used for clinical
routine.
When method is accepted, remember to set replicate to 1 as default for the
unknown samples and to set replicate of controls as desired.
In order to maintain the instrument's performance, please follow instructions in
chapter 11, Maintenance.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...