Semiannual Maintenance
6.9.Maintenance
8.Semiannual Maintenance?
6.9.8.Semiannual Maintenance
REPLACE BOTTLE FILTERS/DRAINAGE FILTER
6.9.Maintenance
1.Daily Maintenance?
The following points are included in the semiannual maintenance:
• Replacement of the Wash and Rinse tubings
• Replacement of the Drainage Filter.
1.
In Maintenance mode
press the number
8
key.
2.
Press the
yes
key.
3.
When the filters are re-
placed press the
enter
key to
confirm this step.
Replace Wash and Rinse Tubings
In each of the Washing Solution and Rinse Solution Bottles there are tubings with
bottle filters at the end that prevent particles from entering valves. It is advisable
to replace these tubings and filters once every six months.
Disconnect the Washing Solution and the Rinse Solution Bottles by using the
quick release. Unscrew the caps from the bottles.
Unscrew the nut under each cap and pull the tubing from the nipple.
Push a new tubing on firmly and screw the nut back in position.
Screw the caps back on the bottles before you connect them to the instrument.
WARNING! Washing Solution may cause
sensitization by skin contact. Wear protective
gloves!
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...