Maintenance
Phadia 100 - User Manual
P a g e 1 1 . 2 4
12-3501-20/12
6.9.4.System Rinse
A completed assay run ends with a System Rinse automatically. If the process
have been aborted by a processing error, you have to perform
6.9.4.System
Rinse
.
Make sure that there is enough Rinse and space in the Waste Bottle (it normally
should if it was enough at the start of the assay run).
1.
In Maintenance mode
press the number
4
key.
2.
Press the
yes
key.
3.
Press the
enter
key to con-
firm this step.
When the System Rinse is done finish off as after a completed assay run.
6.9.Maintenance
1.Daily Maintenance?
6.9.Maintenance
4.System Rinse?
6.9.4.System Rinse
PRESS ENTER TO START PROCESS