Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.61
12
36
36
36
36
36 Check if Washing Solution is loaded
Shake Washing Solution bottle gently. The bottle with Washing Solution will
foam. Make sure bottle connection is ok.
37
37
37
37
37 Check that Rinse Solution is loaded in Rinse bottle
and Washing Solution is loaded in Washing Solution Bottle
Shake both Washing and Rinse Solution bottle, the bottle with Washing Solution
will foam.
38
38
38
38
38 Check mix-up between calibrators
Check reagents used before unloading from instrument. Print Distribution List if
not already printed. Check that right reagents are in assigned position according
to Distribution List. Always compare loaded run to Distribution List.
39
39
39
39
39 Check serum sample for IgA deficiency
If a "Low RU" message occurs in an EliA IgA test, this is a strong hint for IgA
deficiency, which is relatively frequent in celiac disease patients. In such samples
the total IgA level should be determined and EliA Celikey IgG and/or EliA
Gliadin IgG should be tested, if celiac disease is suspected.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...