TEMPERATURE SENSORS DIFFER
What can cause the problem:
There is a difference in temperature between the sensors in the upper and lower
parts of the instrument.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Check if the fan in the top box is running. Check by holding the hand on the back
of the instrument upper left side. If no air stream the fan has stopped.
If the error remains contact your local Phadia representative.
CHECK CAPS ON VIALS
There is a vial cap left on one or more of the vials in the Sample Carousel.
Make sure that all vial caps are removed. This message is inactivated by default.
If the error remains contact your local Phadia representative.
WASH/RINSE CHECK NOT OK
What can cause the problem:
The distribution of Washing Solution is insufficiant. Washing Solution might not
have been loaded. This error will also occur if either Rinse Solution or Washing
Solution have been loaded in both bottles. The error can also occur when the
Washing Solution has not been mixed thoroughly.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Make sure that the tubing from the Washing and Rinse Solution bottles are not
twisted and that the tubings are properly connected to the bottles. To avoid
twisted tubes, connect tubings after Washing and Rinse Solution bottles have
been recapped. Check that the tubings to the bottle filters are connected
properly. For additional information see Chapter 11, Semiannual Maintenance.
Check that Rinse Solution is loaded in Rinse bottle and Washing Solution is
loaded in Washing Solution Bottle. Shake both Washing and Rinse Solution
bottle, the bottle with Washing Solution will foam.
Always make sure to mix the Washing Solution according to the Directions for
Use for Washing Solution.
If the error remains contact your local Phadia representative.
WASH/RINSE HAVE BEEN MIXED
What can cause the problem:
Washing Solution and Rinse Solution bottles or connections have been mixed or
Rinse Solution could be contaminated.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...