Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 2 2
12-3501-20/12
FATAL #16.03 PIPETTE PUMP MOTOR CAN'T FIND FLAG
What can cause the problem:
The Pipette Pump motor does not seem to work.
Corrective actions:
Contact your local Phadia representative.
FATAL #16.04 FLUOROMETER MOTOR CAN'T FIND FLAG
What can cause the problem:
The Fluorometer motor does not seem to work.
Corrective actions:
Contact your local Phadia representative.
FATAL #16.05 CAP CAROUSEL MOTOR CAN'T FIND FLAG
What can cause the problem:
The ImmunoCAP Carousel motor does not seem to work. Make sure that nothing
is blocking the movement of the ImmunoCAP Carousel.
Corrective actions:
For Phadia 100 Ver.3.0, use the menu
6.13.Shutdown
and follow the instruc-
tions displayed to safely shut down the instrument. Switch Phadia 100 off, wait
10 seconds and switch on again.
Follow instructions in Action List in order to identify or eliminate the errors.
Check/Remove if the movement of the ImmunoCAP Carousel has been ob-
structed. Check the processing chamber for loose ImmunoCAP/EliA Wells, look
under the Elution Wells Disc, check also under the Wash Rake. Clean the
processing chamber. If problem persist call service.
Rerun the results that have not been reported. Check that sample volume is
enough including the sample tube dead volume, otherwise new patient sample
is needed. Use
1.1.Enter requests
and put in the requests that have not been
reported.
If the error remains contact your local Phadia representative.
FATAL #16.06 SAMPLE CAROUSEL MOTOR CAN'T FIND FLAG
What can cause the problem:
The Sample Carousel motor does not seem to work. Make sure that the Sample
Carousel is placed in its right position and that nothing is blocking the
movement.
Corrective actions:
For Phadia 100 Ver.3.0, use the menu
6.13.Shutdown
and follow the instruc-
tions displayed to safely shut down the instrument. Switch Phadia 100 off, wait
10 seconds and switch on again.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...