Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 5 0
12-3501-20/12
REAGENT BLANK IS TOO HIGH
If the Reagent blank is >187.5 RU this error message will occur.
What can cause the problem:
This error might be caused by contamination of the Development Solution or the
Elution Wells Disc might need cleaning.
Normal values
For
ImmunoCAP
between 10 - 20 RU and for
EliA
between 15 - 35 RU.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
18, 19, 31.
If the error remains contact your local Phadia representative.
RINSE BLANK IS TOO HIGH
If the Rinse blank is > 31.25 RU this error will occur.
What can cause the problem:
This error might be caused by contamination of the Rinse Solution or that the
Rinse Bottle has been filled with the wrong solution. Dirty cuvette or Elution
Wells Disc can also cause the error.
Normal values
5 - 20 RU
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Perform Monthly Maintenance as corrective action before next run in order to
clean the instrument. Monthly Maintenance should be done even if the regular
Monthly Maintenance has been done recently. Something has happened during
the run and the instrument needs to be cleaned, like after Rinse Solution and
Washing Solution distribution problems, lack of maintenance or vacuum
problems. After going trough all other actions on the list first and thereby
eliminate the error it could be that Monthly Maintenance can be skipped this
time.
If the error remains contact your local Phadia representative.
THE DIFFERENCE BETWEEN REAGENT BLANK
AND RINSE BLANK IS TOO LOW
What can cause the problem:
Broken cuvette. Rinse Solution or Washing Solution in both bottles. Rinse
Solution that is contaminated.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number,
11,
26
.
If the error remains contact your local Phadia representative.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...