Error Handling
12-3501-20/12
Page 12.19
12
FATAL #03.00 SPURIOUS NMI INTERRUPT
What can cause the problem:
This error is caused by an external disturbance or it has occurred an error in the
electronics. The assay run is aborted.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Empty instrument manually and run Monthly Maintenance. Unload Sample
Carousel, open Processing lid and lift up ImmunoCAP holder disc. Throw away
ImmunoCAP or EliA Wells. Check that no ImmunoCAP or EliA Wells are left
in or under the Elution Wells Disc. Check also under Wash Rake. After
emptying the instrument, perform a complete Monthly Maintenance, see
Chapter 11.
Load old run and rerun Samples. Check that sample and reagent volumes are
enough for the rerun. Look at consumption list and include dead volume for the
sample tubes. Rerun the results that have not been reported. Load the assay run
again by first using
1.5.Load Old Run.
If the error remains contact your local Phadia representative.
FATAL #04.00 CONTROLLER HALTED
What can cause the problem:
This error is caused by an external disturbance or it has occurred an error in the
electronics.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Empty instrument manually and run Monthly Maintenance. Unload Sample
Carousel, open Processing lid and lift up ImmunoCAP holder disc. Throw away
ImmunoCAP or EliA Wells. Check that no ImmunoCAP or EliA Wells are left
in or under the Elution Wells Disc. Check also under Wash Rake. After
emptying the instrument, perform a complete Monthly Maintenance, see
Chapter 11.
Load old run and rerun Samples. Check that sample and reagent volumes are
enough for the rerun. Look at consumption list and include dead volume for the
sample tubes. Rerun the results that have not been reported. Load the assay run
again by first using
1.5.Load Old Run.
If the error remains contact your local Phadia representative.
WARN #05.00 POWER TEMPORARY LOST
A short power down (<10 msec) has occurred. The processing has not been
disturbed.
What can cause the problem:
If this error occurs frequently it is an indication of a bad mains supply, or that
something is wrong in the power supply unit.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...