Error Handling
Phadia 100 - User Manual
P a g e 1 2 . 3 2
12-3501-20/12
FATAL #40.00 Position:
[nn]
DILUENT VIAL/TUBE EMPTY
What can cause the problem:
The level detector has detected fluid before aspiration but not afterwards when
it is leaving the diluent vial or tube. This warning indicates that the diluent vial
or tube may be out of fluid. The pipette might not have aspirated correct volume.
There might have been an air bubble in the diluent vial or tube which has been
popped when the Pipette tip went through it.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Rerun the results that have not been reported. Check that sample volume is
enough including the sample tube dead volume, otherwise new patient sample
is needed. Use
1.1.Enter requests
and put in the requests that have not been
reported.
If the error remains contact your local Phadia representative.
#41.00 Position:
[nn]
CONJUGATE VIAL EMPTY
What can cause the problem:
The level detector has detected fluid before aspiration but not afterwards when
it is leaving the conjugate vial. This warning indicates that the conjugate vial
may be out of fluid. The pipette might not have aspirated correct volume. There
might have been an air bubble in the conjugate vial which has been popped when
the Pipette tip went through it.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Rerun the results that have not been reported. Check that sample volume is
enough including the sample tube dead volume, otherwise new patient sample
is needed. Use
1.1.Enter requests
and put in the requests that have not been
reported.
If the error remains contact your local Phadia representative.
#42.00 Position:
[nn]
DEVSOL VIAL EMPTY
What can cause the problem:
The level detector has detected fluid before aspiration but not afterwards when
it is leaving the development solution vial. This warning indicates that the
development solution vial may be out of fluid. The pipette might not have
aspirated correct volume. There might have been an air bubble in the develop-
ment solution vial which has been popped when the Pipette tip went through it.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors.
Rerun the results that have not been reported. Check that sample volume is
enough including the sample tube dead volume, otherwise new patient sample
is needed. Use
1.1.Enter requests
and put in the requests that have not been
reported.
If the error remains contact your local Phadia representative.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...