Assay Run Data Evaluation
Phadia 100 - User Manual
P a g e 5 . 1 4
12-3501-20/12
Messages that can occur
MEDIA IS FOR OTHER INSTRUMENT
The USB memory, external USB hard drive or diskette is labeled with another
instrument Id. Data can only be exported to media labeled with the same
instrument Id.
MEDIA ERROR, ABORT?
The USB memory, external USB hard drive or diskette is not formatted or it is
damaged. Format the media or change it. Press
no
to continue or press
yes
to
abort.
MEDIA FULL, ABORT?
The USB memory, external USB hard drive or diskette is full. Delete old files or
change media. Press
no
to overwrite data or press
yes
to abort.
DATA ON MEDIA, OVERWRITE?
Data is already stored on the USB memory, external USB hard drive or diskette.
Press
yes
to overwrite data or press
no
to change the USB memory, external
USB hard drive or diskette.
MEDIA NOT INSERTED, ABORT?
Insert a USB memory or a diskette, or connect an external USB hard drive. Press
no
to continue or press
yes
to abort.
MEDIA WRITE-PROTECTED, ABORT?
Remove the write protection or change the USB memory, external USB hard
drive or diskette. Press
no
to continue or press
yes
to abort.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...