Phadia 100 - User Manual
P a g e 9 . 1 0
12-3501-20/12
** This menu can not be reached for the default methods.
5.2.3.2.6.Low RU on zero-point**
You can enable or disable this function. Disable for example when you run EliA
methods in order to avoid Low RU error.
5.2.3.2.7.Quotient Point
Define the default calibrator point to be used for calculating the quotient when
reporting results. Confirm by pressing
enter.
5.2.3.2.8.Add Point**
Used to add a calibrator to the selected method.
Define identity, concentration**, replicate**, normal response, expected s,
limit s upper RU, limit s lower RU, limit s precision and limit s deviation.
5.2.3.2.9.Edit Point**
Here you can change some parameters for the calibrators. Select which
calibrator, confirm with
yes
.
You can change concentration**, replicate**, normal response, expected s,
limit s upper RU, limit s lower RU, limit s precision and limit s deviation.
5.2.3.2.10.Delete Point**
Select calibrator point to be deleted. Confirm with
yes
.
You will be prompted to
PRESS ENTER TO DELETE POINT.
Press
enter
and
POINT DELETED
is shown on the display.
5.2.3.3.Blank**
Select if the Blank should be subtracted or not.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...