Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.41
12
ImmunoCAP Specific IgG, ImmunoCAP Specific IgG4
and
EliA
, contamination
problems.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
7, 22, 24, 38.
If the error remains contact your local Phadia representative.
CALIBRATION CURVE OLDER THAN RECOMMENDED
What can cause the problem:
The Calibration Curve is too old.
Corrective actions:
Run a new Calibration Curve in the next assay run.
Curve Control Errors
Curve Control Errors occur when Curve Controls are outside the inner limits of
the Calibration Curve.
If the Curve Controls are outside the inner limits, you get HIGH*, HIGH, LOW*
or LOW under Curve Controls on the Laboratory Report.