Methods, Test Panels, Article Management
Phadia 100 - User Manual
P a g e 9 . 3 4
12-3501-20/12
6.6.Article Management
ImmunoCAP 100 Ver.3.0
14:10:17
1.Prepare Run?
ImmunoCAP 100 Ver.3.0
14:10:17
6.Install/Service?
6.6.1.Add Article
If you are working with a bar code reader and you receive a new article, for
example a new allergen you need to add the article package code to the
instrument. You must be logged in with laboratory password if you want to work
with articles.
1.
In Prepare Run mode se-
lect
6
.
2.
Press
yes
.
3.
In Install/Service mode
select
6
.
4.
In Article Management
mode select
yes
.
5.
Press
yes
.
6.
Type the new article ID
(up to three characters) and
press
enter
.
7.
Press
no
until the option
you want appears and press
yes
.
8.
Type the article method
(up to four characters) and press
enter
to confirm this step.
9.
Type the article text (up
to 16 characters) and confirm
this step with
enter
.
6.Install/Service
1.Login/Logout?
6.6.Article Management
1.Add Article?
6.6.1.Add Article
Article ID: 123
6.6.1.Add Article
Article type: UNDEFINED?
6.6.1.Add Article
Article Method:_
6.6.1.Add Article
Article Text:_
6.Install/Service
6.Article Management?
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...