Phadia 100 - User Manual
Assay Run Editing Functions
12-3501-20/12
Page 4.13
3
4
5
6
4
1.4.1.2.Change Test
[1239]
Enter Replicate Number:1_
1.4.1.2.Change Test
[1239]
Test Dilution:INSTRUMENT DILUTION?
1.4.1.2.Change Test
[1239]
Test Dilution:PREDILUTED?
1.4.1.2.Change Test
[1239]
Enter Dilution factor:5_
11.
Type the desired number
(between 1 and 5) of replicates
for the selected test and press
enter
.
If you do not want to change
the replicate just press
enter
.
12.
Press
yes
if you want the
instrument to perform the dilu-
tion otherwise
no
.
13.
Press
yes
if the sample is
diluted manually before placed
in the instrument.
14.
Type the new factor over
the old. Press
enter
.
If the sample is to be diluted in the instrument you can select factor 1, 2, 5, 10,
20, 50 or 100.
If the sample is prediluted you can select a factor between 1 and 1000.
Messages that can occur
NO ACTIVE CALIBRATORS, ADD TO RUN?
There is no stored calibration curve for the method.
NO TEST POS. LEFT. Skip Test?
There is no position left for the ImmunoCAP/EliA Well. If you press
yes
the test
will be removed.
NO SAMPLE POS. LEFT. Skip Test?
There is no position left for sample. If you press
yes
the test will be removed.
PREDILUTED?
Press
yes
if the test is prediluted ten times (this example).
INSTRUMENT DILUTION?
Press
yes
if the test is to be diluted ten times (this example). The dilution is
carried out by the instrument.
TEST ALREADY ENTERED
The test name has already been entered for the sample.
TOO MANY METHODS AT THE SAME TIME
The number of methods possible to enter was exceeded. Maximal number of
methods that can be entered in one run is four (4).
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...