Routine Assay Run
Phadia 100 - User Manual
P a g e 2 . 3 8
12-3501-20/12
Enter Request
You enter a request by starting with the sample identity and then adding the test
identities. You can enter the identities from the keyboard or with a barcode
reader. If a calibration curve is to be included in the run, you specify the
calibration curve as described in the Calibration Curve Assay.
You will be prompted to select whether or not Quality Controls should be
included with the specified method if Quality Controls are given the status
Manual. Quality Controls with status Active will be added automatically to the
assay run. Status of Quality Controls are set in
5.2.9.1.Add QC
.
The method used is the one selected as default (menu level
6.4.2.Default
Method
) when the program was installed. You can change the method by
pressing
alt
when entering the test identities.
If you enter more tests then there are available positions, you will be asked if you
want to skip the whole sample identity or just the tests left for that sample
identity. You can also leave the Enter Request mode before all the positions are
filled by pressing
enter
. If you press
mode
, the request will not be saved.
The number of ImmunoCAP/EliA Wells that can be entered in one assay run is
48. This is also the maximum number of samples that can be placed in the sample
carousel.
1.
Press
yes
.
2.
Press
yes
.
Warnings that may prevent entering
REMOTE REQUESTS ENTERED
The currently prepared run has been prepared on a remote PC and cannot be
edited in the instrument.
RUN IS LOADED
Loading of reagents, ImmunoCAP/EliA Wells or samples has already been
started and the run cannot be changed.
PC NOT READY
Instrument is set in PC remote mode, but the PC is not set in "communication
mode".
NO ROOM FOR MORE REQUESTS
The currently prepared run occupies all ImmunoCAP/EliA Wells, sample or
reagent positions.
Phadia 100 Ver.3.0 14:10:17
1.Prepare Run?
1.Prepare Run
1.Enter Request?
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...