Assay Run Editing Functions
Phadia 100 - User Manual
P a g e 4 . 3 4
12-3501-20/12
3.
Choose the type of QC
you want, CONCENTRA-
TION, QUOTIENT or
QUALITATIVE, by pressing
no
until the desired type is dis-
played. Then press
yes
.
Order
The following step describes how you can change when in the assay run the
Quality Control will be processed.
1.
In Edit QC mode press
16
.
2.
Press
yes
.
3.
Delete or type over the
old order number. Type a num-
ber between 1-99. A low num-
ber means that the Quality
Control will be processed be-
fore Quality Controls with
higher numbers. Type 0 (zero)
to use the default order of the
Quality Control. Press
enter
to confirm this step.
NOTE!
Only use this function if you want to change the default order of the
Quality Controls.
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
16.Order?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Order: 0
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Type of QC: CONCENTRATION?
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...