Routine Assay Run
Phadia 100 - User Manual
P a g e 2 . 3 2
12-3501-20/12
Principles of Operation
The instrument is controlled from the panel at the front of the instrument.
You load the reagents and samples in the Sample Carousel to the left of the
instrument. Positions are marked to simplify the loading of the reagents.
You load ImmunoCAP/EliA Well carriers at the top of the middle section of the
instrument.
Each required reagent, sample or ImmunoCAP/EliA Well will be prompted on
the display in the Load and Start mode. You load it and confirm the step with
the instrument.
The Washing Solution, Rinse Solution and Waste bottles are placed just behind
the carousel to the left of the instrument.
The instrument is also equipped with a printer and a diskette drive. You find the
printer to the right in the middle section and the diskette drive at the right on
top of the instrument.
Workflow
Enter Request
Go to Prepare Run mode and select
1.1.Enter Request
. Type
cal
and choose
method. If not all methods are displayed, you can find them if you press
alt
.
If
a calibration curve is not requested, type the sample identities, choose test
(allergen/autoimmunity antigen) respectively method. Curve Controls will
automatically be included in the assay setup, if calibrators are not run.
Controls defined in the method as Active will automatically be added to the
assay setup when the first patient request is entered for that method.
Print List
Select
1.2.Print Lists
to print out a wanted list:
• request list to check if request is correct
• consumption list to see the required volume of reagents and samples
• distribution list to know where samples and reagents are placed, in case you
load the wrong sample or reagent and want to change manually.
• ImmunoCAP/EliA Wells are not marked with short names so that manual
placement of ImmunoCAP/EliA Wells is not recommended.
For further and detailed information of all printouts, see chapter 10, Printouts.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...