Phadia 100 - User Manual
Assay Run Editing Functions
12-3501-20/12
Page 4.33
3
4
5
6
4
Predilution factor
The following step describes how you can change the predilution factor.
1.
In Edit QC mode press
13
.
2.
Press
yes
.
3.
Delete or type over the
old predilution factor. If the
Quality Control is prediluted
type a dilution factor between
1-100. If the Quality Control is
not prediluted, type 1. Press
enter
to confirm this step.
Instrument dilution factor
The following step describes how you can change the instrument dilution factor.
1.
In Edit QC mode press
14
.
2.
Press
yes
.
3.
Delete or type over the
old instrument dilution factor.
Type the new instrument dilu-
tion factor. Choose between 1,
2, 5, 10, 20, 50 or 100. Press
enter
to confirm this step.
Type of Quality Control
You can set the type of the Quality control to CONCENTRATION, QUO-
TIENT or QUALITATIVE.
1.
In Edit QC mode press
15
.
2.
Press
yes
.
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Dilution factor: 10
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
14.Instr. dilution factor?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
Instr. dilution factor: 10
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
13.Predilution factor?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
1.Status?
5.2.9.2.Edit QC
[sIgE/sIgE Pos]
15.Type of QC?
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...