Phadia 100 - User Manual
Error Handling
12-3501-20/12
Page 12.53
12
If the error remains contact your local Phadia representative.
Low RU, Whole carousel
What can cause the problem:
Cuvette broken. Missed pipetting because no Rinse Solution available or no
Conjugate, Development Solution or Stop Solution loaded. Stop Solution
loaded instead of Washing Solution Concentrate or Washing Solution Additive
instead of Development Solution.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
10, 11, 24.
If the error remains contact your local Phadia representative.
Low RU, Single position
What can cause the problem:
This error might be caused by contamination of the Development Solution or the
Elution Wells Disc might need cleaning. Missed pipetting of Conjugate or
Development Solution.
Missed ImmunoCAP/EliA Well dispensing. The Reagent Blank measured too
high caused by problems with Reagent distribution. For
EliA
missed raw serum
in serum dilution.
Corrective actions:
Follow instructions in Action List in order to identify or eliminate the errors. For
additional information see Explanation of Actions, number
7, 10, 12, 24.
If the error remains contact your local Phadia representative.
FluoroC Errors
FluoroC Errors may occur when you have performed a FluoroC run in order to
check the Fluorometer. If you get NOT OK on the FluoroC Run Report, which
is printed after the FluoroC run is finished, it means that you have a FluoroC
Error. At the end of the FluoroC Run Report you find the Action List with
instructions on how to eliminate the errors.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...