Methods, Test Panels, Article Management
Phadia 100 - User Manual
P a g e 9 . 1 8
12-3501-20/12
5.2.7.2.8.Report quotient
Select report quotient,
YES
,
YES PERCENT
, or
NO
. Confirm with
yes.
Quotient is the ratio between a sample value and the class border.
5.2.7.2.9.Concentration unit**
5.2.7.2.10.Concentration factor**
5.2.7.2.11.Highest concentration**
5.2.7.2.12.Decimals for conc <10**
5.2.7.2.13.Use Lot Specific Code**
Select Lot Specific Code or not with the
yes
or
no
keys. Confirm with
yes
.
5.2.7.2.14.Dilution factor
Type a number between 1 and 999 for
PREDILUTED
or 1, 2, 5, 10, 20, 50 or 100 for
INSTRUMENT DILUTION
Confirm with
enter
.
5.2.7.2.15.Dilution method
Select
INSTRUMENT DILUTION or PREDILUTED
Confirm with
yes
.
5.2.7.2.16.Diluent**
Define the identity of the diluent. Some methods (IgG/IgA) use a special diluent.
5.2.7.2.17.Cut-off concentration*
Type a concentration. Confirm with
yes
.
* This menu can be reached if you are logged in with laboratory password.
** This menu can not be reached for the default methods.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...