Phadia 100 - User Manual
Monthly Calibration
12-3501-20/12
Page 7.5
7
Methods using two curve controls
The following results are possible in the first run for methods using two curve
controls:
CC = Curve Control
1. Both curve
controls are OK.
Results/Curve status
CC1
OK
CC2
OK
Results calculated
Run accepted (OK)
2. One curve
control is OK, the
other is within the
outer limits
OK
OK
HIGH*
LOW*
LOW*
HIGH*
OK
OK
Results calculated
Curve flagged
3. Both curve
controls are within
the outer limits in
the same direction.
LOW*
HIGH*
LOW*
HIGH*
Results calculated
Curve flagged
WARNING: Results must
be judged - based on the
QC results
4. Both curve
controls are within
the outer limits in
the opposite
direction.
LOW*
HIGH*
HIGH*
LOW*
No results calculated
Curve flagged
5. One or both
curve controls are
out of the outer
limits.
OK
OK
LOW
LOW
LOW
LOW
LOW
LOW*
LOW*
HIGH
HIGH
HIGH
HIGH
HIGH
HIGH*
HIGH*
HIGH
LOW
OK
LOW
LOW*
HIGH
HIGH*
LOW
HIGH
OK
LOW
LOW*
HIGH
HIGH*
LOW
HIGH
No results calculated
Curve flagged
NOTE! This symbol * means CHECK:
• The result for one curve control has been
observed outside the specified inner limits
• Indication of change in dose-response com-
pared with the calibration run.
Содержание Phadia 100
Страница 1: ...Phadia 100 English 12 3501 20 12 User Manual...
Страница 81: ...Routine Assay Run Phadia 100 User Manual Page 2 50 12 3501 20 12...
Страница 91: ...Reagent and Sample Handling Phadia 100 User Manual Page 3 10 12 3501 20 12...
Страница 251: ...Printouts Phadia 100 User Manual Page 10 22 12 3501 20 12...
Страница 277: ...Maintenance Phadia 100 User Manual Page 11 26 12 3501 20 12...
Страница 341: ...Error Handling Phadia 100 User Manual Page 12 64 12 3501 20 12...
Страница 348: ...Phadia 100 User Manual Index 12 3501 20 12 Page 15 1 15 Chapter 15 Index...