Operating Instructions | HERAcell® 150i / 240i
13
Annex: Data communication
13
Annex: Data communication
50115191 A / 08.2009
134
13.5.3 Overview of the possible error messages in bit coding
Bit-coded error messages for
temperature and CO
2
General device status, control loop for temperature and CO
2
:
Bit
General device status
0x0002
Device door open too long
0x0004
Display fails to communicate
0x0008
Mainboad parameter implausible (EEPROM faulty)
0x0010
Data logger faulty (device remains operative)
0x0020
Error in disinfection / contraCon
0x0040
No power supply during contraCon
0x0080
Error in autostart
0x0100
Safety circuit activated
0x0200
Bottle turning device fails to communicate
0x2000
Autostart active (info)
0x4000
Disinfection active (info)
0x8000
Device error has occurred (info)
Bit
Temperature control loop error status
0x0001
Sensor breakage
0x0002
Value above set value upper limit
0x0004
Value below set value lower limit
0x0008
Set value implausible
0x0010
Calibration value too high / too low
0x0020
Communication error (to sensor)
0x0040
Communication error (to LM75)
0x0100
Measuring cell parameter implausible (EEPROM faulty)
Bit
CO
2
control loop error status
0x0001
Sensor breakage
0x0002
Value above set value upper limit
0x0004
Value below set value lower limit
0x0010
Calibration value too high / too low
0x0020
Communication error (to sensor)
0x0040
Communication error (to cylinder changeover switch)
0x0080
No gas, cylinders A and B empty
0x0100
Measuring cell parameter implausible (EEPROM faulty)
0x0200
Cylinder A empty
0x0400
Cylinder B empty