Operating Instructions | HERAcell® 150i / 240i
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
89
50115191 A / 08.2009
Quick access to bottle turning device
levels:
To directly access the Bottle turning device
submenu, touch the Bottle turning device
icon in the main menu.
Switching the O
2
control on and off:
Depending on the requirements to the work process,
the O
2
control can be switched on and off. This setting
is possible only on versions with the optional O
2
/N
2
control.
1. To toggle between two states of the O
2
control:
>
Press the
+
key [4].
or
>
Press the
-
key [2].
2. The value change appears in the display [1].
3. To accept and save the setting:
>
Press the
E
NTER
key [3].
>
The system returns to the Options menu.
O
2
value display:
If the O
2
control has been switched off, the
O
2
display does not show an actual value (-
- -).
This procedure provides protection for the
O
2
sensor.
If the set value is set to 21%, the O
2
control
loop is not monitored. This applies to both
O
2
control ranges:
• Control range I: 1% to 21%
• Control range II: 5% to 90%
In this case, the O
2
display shows the
actual value.
Airing the work space:
If the device was operated with O
2
or N
2
,
the work space must be aired after the O
2
control has been switched off.
Fig. 58: Switching the O2 control on and off