Operating Instructions | HERAcell® 150i / 240i
99
50115191 A / 08.2009
8
Shut-down
8.1 Shutting the device down
Contamination hazard!
If the work space surfaces are contaminated, germs can be
transferred to the environment of the device.
In case of a shut-down, the device must be decontaminated!
1. Remove the culture containers and all accessories from the work
space.
2. Pump the water out off the water reservoir (see chap. 9).
3. Fill in 350 ml water and start the contra- con decontamination routine.
4. Turn the device off using the power switch when the contra-con
decontamination routine has been completed.
5. Unplug the power connector and protect it against accidental
reconnection.
6. Close the CO
2
/O
2
/N
2
supply system shut-off valves.
7. Disconnect the gas pressure hoses from the sleeve at the rear of the
device.
8. Clean the work space and wipe the device dry.
9. Until the device is shut down, the work space must be continuously
ventilated. Leave the glass door and the outer door open and secure
them in this state.