Operating Instructions | HERAcell® 150i / 240i
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
7
Handling and control (iCan™ touchscreen controller)
72
50115191 A / 08.2009
7.8 Setting the O
2
set value
This setting is possible only on versions with the
optional O
2
/N
2
control.
1. Press the O
2
DISPLAY
key [1].
>
The O
2
menu is displayed.
2. To exit the O
2
menu:
>
Press the
E
ND
key [2].
3. To set the O
2
set value:
>
Press the
S
ET
VALUE
key [3].
To increase the set value:
>
Press the
+
key [6].
To reduce the set value:
>
Press the
-
key [4].
4. To accept and save the set value:
>
Press the
E
NTER
key [5].
>
The system returns to the main menu. The O
2
displays shows the actual value currently
measured in the work space.
Factory settings:
Depending on the type of the O
2
sensor,
one of the two O
2
control ranges has been
preset at the factory:
Control range I: 1% to 21%
Control range II: 5% to 90%
Usage of the process gases:
For set values below 21% O
2
, the device
must be connected to a nitrogen supply
system.
For set values above 21% O
2
, the device
must be connected to an oxygen supply
system.
Since the O2 concentration of the air is
nearly 21%, the O2 control and the control
loop error monitoring system are disabled
at a set value of 21%.
The control loop error monitoring system
(optional) is enabled.
Fig. 38: Setting the
O
2 set value
Содержание Heracell 150i
Страница 156: ......